EMS Technical Data Sheets

Retriever Operation Manual

EMS Catalog #62706

Introduction

What is the Retriever

2100-Retriever, or, from now on simply Retriever, is a bench-top model for thermally processing slides of the formalin-fixed, paraffin embedded tissues prior to immunostaining. The model is designed to ensure identical processing of all the samples during a processing cycle, as well as the iidentical processing of the samples between individual sessions.

How Does It Work

The Retriever is developed on the basis of an autoclave. There have been earlier reports that autoclaving formalin-fixed sections may be in no way lless effective than processing them in a microwave. After a year of testing we found the optimal configuration of such an autoclave. The core principleis the heating of the chambers with the slides at high temperature (>120 degree C) and high pressure. The chambers reach the required temperatureand remain there for a fixed time. The specially designed thermal walls of the unit control the cooling rate of the inner chamber and slides. The unit is fully automatic.

Advantages Over Microwave

Reproducibility. There is no boiling of the buffer, and no cold zone effect, usual to a microwave, when some of the sections are heated less than others. All sections are processed uniformly and always for the same length of time. There are no variations from cycle to cycle so you can compare the sections processed with a gap of a couple of months (or years). Since the cycle is fully automatic, the heating of the sections, exposure to high pressure and temperature, and even cooling (also quite important) are standardized. So are your results.years). Since the cycle is fully automatic, the heating of the sections, exposureto high pressure and temperature, and even cooling (also quite important) are standardized. So are your results.

Quality of the results. The Retriever preserves the tissue morphology, thus your images will be of much higher quality. For some antigens we found that the intensity of the staining is much higher than with standard methods (especially for nuclear antigens, such as estrogen receptor).

Your time. All you have to do is to put slides in a chamber and push the button. While the unit is working, you do not have to be nearby, you can carry on with other matters.

All antibodies that can be used on microwave treated tissues can be also used in Retriever.

Retriever and Acessories (Scheme)

The supplier Retriever set contains the processing unit, slide chamber (allows to fit 18 Slides) and Chamber Rack for 6 Slide Chambers (108 slides in total). On the control panel the Start Button and indicator lights are located that show different stages of the processing program.

Please get acquainted with the location of various parts of Retriever and the signals displayed by indicator lights,

A Depressurization Valve  
B Display Panel  
C Pressure Rize Indicator  
D Body of Retriever  
E Slide Chamber  
F Chamber Rack  
G Display Lights  
  L1- Power on - Green
  L2- Heating on - Orange
  L3 - The cycle conditions are reached Yellow
  L4- Cycle complete Green
  L5 - Fault Light Red

Getting Started

Retriever is an autoclave and thus has to be handled appropriately and with care.

Before using the Retriever for the first time please take time to read the following pages to familiarize yourself with the operation of the unit. We strongly recommend that all users of the Retriever are trained in its operation.

Place the Retriever on the bench-top at the open space since some vapor will be coming out of the pressure indicator. Please use the appropriate voltage outlet if your laboratory has both 110-115V and 220V. All models for US and Japan are standardly shipped as 110V models, unless specifically requested otherwise.

WARNING! Mains outlet MUST BE EARTHED (GROUNDED)

The mains plug should always be easily accessible as it is to be relied upon as "the means of disconnection".

Please notice the zones (as indicated above) of the heat coming out of the unit. When cycle is run, the Body may be hot. Some hot steam may be released from the site of Pressure indicator and Depressurization Valve.

WARNING! The Depressurization Valve must be closed before the unit is started for the cycle running!

Fixing Tissue Samples

We have developed the Retriever for formalin fixed, paraffin embedded tissues, the most routine material available from tissue collections and easy to prepare from fresh tissue samples. A routine protocol used in the majority of pathology laboratories is sufficient. Your approximate protocol should look like this:

Protocol R1

Notes

When you are preparing the specimen yourself, please wash all residual fixative properly by extended incubation in 50% and 70% ethanol prior to submitting it to further dehydration and embedding to paraffin.

Isolating tissue specimen please avoid tissue autolysis. To prevent this the best solution is to place the tissues straight to the cold fixative. Alternatively use specialized solutions for transporting tissues. Use a cold plate to handle the tissues whenever possible.

Consider that very hard tissues may require an extended (up to 48 hours) fixation time. 

To preserve good infrastructure of certain tissues such as lung, pancreas, liver, a fixative perfusion is required. Although this may not be necessary for most immunohistology stainings, it may be desirable in cases when ultrastructure of a particular organ area is of importance. If the tissues have to be compared with respect to expression of a certain antigen, please control that they all undergo identical fixation (which includes size of samples, time of incubation).

Cutting Tissue Sections

Protocol R2

Notes

Important. Use specialized coated tissue slides for sections that will further be processed for antigen unmasking. The best results will be obtained using Superfrost Plus or Gold or Polylysine-coated Slides from Menzel-Glazer. Alternatively, you may prepare your own coated slides (see the protocol below).

Important. When the slides are cut and are dried/spread overnight, use them the same day. Otherwise, place slides in a tightly closed slide box and store at 2-4°C. This prevents slow degradation of many molecules in your sections. According to our experience, storage at room temperature may dramatically affect immunostaining for some antigens already within 1st week.

Making Apes Slides

Protocol R3

The APES-solution should always be freshly made. 1L solution is sufficient for the treatment of 400 microscope glasses.

Dewaxing Tissue Sections

Protocol R4

Choosing the Buffer

Two important issues determine the choice of the buffer:

If your laboratory has good experience with a routine Citrate (pH6) and EDTA (pH 8 or 9) buffers you may continue using those. We strongly advise against using with Retriever Tris-based buffers, including the “ Universal” buffers supplied by some companies .

For best results we recommend using concentrated buffers available in the Retriever section of our catalog:

R-Buffer A pH 6.0 (in place of Citrate 6)

R-Buffer B pH 8.0 (in place of EDTA pH 8 or 9)

R-Buffer C pH 4.5 (very rarely used buffer. Helps to retrieve some antigens)

R-Buffer U pH 6.0 (usually the best buffer to start with when you test antibodies developed against a peptide epitope).

For most routine formalin, formalin-Zn, acid formalin tissues these buffers are optimal with respect to quality of antigen retrieval and morphology preservation.

However, if your routine procedure of fixation is gentle, or you use more gentle than routine formalin fixatives, we recommend using the G variants (“G” stands for Gentle) of the same buffers.

The choice of the buffer also depends on the nature of the antigen. We advise you to first run a test for the most appropriate buffer using the tissues where the expression of the antigen does occur.

General guidelines for choosing the buffer:

Most of the nuclear antigens (apoptosis-related, survival-related, proliferation-related, transcription factors) - R-buffer A.

Cell adhesion molecules, cell membrane antigens (extracellular domain) R-buffer A, rarely buffer B

Cytoskeleton and cytoskeleton-associated molecules - Buffer A Intracellular domain of some adhesion molecules and surface receptors -

Buffer B or C

Most of antibodies raised against the linear peptide - Buffer A or buffer U

Processing Sections in Retriever

Protocol R5

Important. We generally advise you to start the tissue processing procedure in the evening of the day preceding the planned immunostaining experiments. For the best results we advise to allow the Retriever to run the cycle and leave the sections to cool in buffer overnight (over 10 instead of 2 hours).

For many routine antigens the processing may be shortened. In that case leave the sections for 30 minutes after the processing cycle stops, then open the Retriever by first de-pressurizing the unit by opening the pressure valve (beware of the steam) and proceed further as described above. However, please check on several control slides for particular antibodies you are using that this procedure is suitable for a particular tissue material and antibodies that are being used in your laboratory.

Processing Sections

Slide Chamber. Up to 18 slides fits one chamber. Place the slides in the same orientation of the section side
Place Slide Chambers into the Rack and pour the processing buffer into the Chambers to fill all free space.
Fill the Body of Retriever with 750 ml of deionized water and place the Rack into Retriever.

Close the lid as shown. Close the Valve. Push the Start button on the display. The light indicators (shown on the left) will be illuminated as the cycle proceeds.

The Cycle will run for approximately 20 minutes. The following sequence of events occurs:

L1 illuminates green
L2 illuminates orange

As the temperature rises, air will be displaced by steam through the Air Bleed Device located in the lid. Pressure Indicator ( ) will rise indicating the unit is now pressurized.

The Temperature and Pressure conditions are reached when:
Lights:

L1 illuminates green
L2 flashes orange
L3 illuminates yellow

Cycle is completed when the Buzzer sounds.

Lights: L1 illuminates green
L4 illuminates green

(Note: L4 remains green until a new cycle is started or the unit is disconnected from the mains power)

Immunostaining

Protocol R6

After processing in Retriever use the sections within the same day

Important: when choosing the detection system for staining processed sections, aim for a sensitive one. This should ensure a quality result in the end.

Examples of Staining

Examples of Processed Tissues

formalin fixed tissue microwaved tissue antigen retriever tissue

This is a section of intestinal tissue fixed with formalin and embedded into paraffin.

Processing the tissue in microwave-type machines results in damage of the morphology of the gentle tissues.

No damage to morphology occurs, however, after processing the tissue in Retriever, although the unmasking is usually better than after the microwave.

Compare these three sections of squamous epithelium: Section A was untreated, section B was processed in a microwave, and section C - in Retriever. Note the integrity of the tissue after Retriever, while it is obviously damaged by microwave.

Here we show several examples of immunostaining of formalin fixed, paraffing-embedded tissue sections processed in Retriever. Standard processing cycle in buffer A (sections E,F,G) or buffer B (D) was used. The antibodies form were used against CD8 (D), E-cadherin (E), Ki-67 (F) and PCNA (G). APES coated slides were used.

This is a standard quality of morphology and staining you must expect. We have chosen these antigens since they are widely used as markers and the intensity and quality of staining may be evaluated by most of potential users. Please note the quality of the tissue morphology in all preparations presented.

Continued Operation

To ensure your Retriever gives you the years of service for which it was designed, it is important to remember a few "do's" and "don'ts" with regard to the operation of the unit and to carry out the simple care and maintenance procedures on a weekly basis.

Do Ensure That

  1. You read these instructions and always follow the operating sequence.
  2. The work surface on which you will place the Retriever is flat and solid.
  3. If you put in racks/staining jars other than the supplied with the Retriever, the material the former is made of can withstand the temperature and pressure
  4. Tthe water level is maintained regularly with clean distilled or deionised water only.
  5. The unit is in a "draught free" environment and is positioned not less than 250 mm from adjacent walls.
  6. You only use green sealing gasket (219500) and that it is changed at the end of it's life, if visibly damaged, or when shrinkage has occurred, see "Fault mode - 5".
  7. The lid is securely closed when the unit is not in use, to avoid the risk of accidental damage. Never leave in position as shown in F1.
  8. You quote your model details, serial number and date of purchase when contacting PickCell Laboratories or your supplier.

Do Not! ..

  1. Touch the unit whilst in operation - it gets HOT
  2. Attempt to remove the lid during operation. DO NOT LUBRICATE GASKET
  3. Lose this operating instruction manual. If lost, please download it at www.antigenretriever.com
  4. Add any chemicals whatsoever to the water.
  5. Attempt to use in buffers volatile substances, toxic materials, or inappropriate loads such as easy melting plastic.
  6. Place the unit on heat sensitive surfaces, polished wood or glass.
  7. Open the Depressurization Valve (A) during the cycle.
  8. Lleave the Depressurization Valve (A) in the "open" position when placing the lid upside down on a work surface.
  9. Immerse the unit or electrical cord in water when cleaning.
  10. Use abrasive materials or lubricants when cleaning.
  11. Drop or abuse the unit.
  12. Use in areas of risk associated with flammable materials or gasses.
  13. Attempt to change fuses until the unit has been unplugged from the mains. Only qualified persons should change fuses.
  14. Reach over the unit when removing cover, to do so may cause burns from rising heat and steam.
  15. Press start button once cycle has been started as this will reset the cycle timer to zero.

Care and Maintanance

Cleaning Retriever

WARNING! Disconnect the Retriever from the main’s power supply be fore cleaning.

Please always use only deinoized water to fill the internal chamber of retriever. In this case no scaling occurs. If some scaling occurs due to buffer spilled, rinse the inside of the Retriever with light de-scaling agen (like de-scaling liquid for electric cookers and kettles). Rinse with distilled water and wipe dry.

If the unit gets dirty, clean both interior and exterior with warm, soapy water ensuring the electrical parts are kept dry.

The sealing gasket in the lid of Retriever must be cleaned if any leakag occurs (the vapor coming from the side of the lid, not through the pressure indicator, the unit has problems with keeping the pressure during the cycle). To perform this operation:

Green Sealing Gasket

  1. Remove the Gasket from inside the lid and clean with warm, soapywater.
  2. Rinse thoroughly, shake dry, do not wipe.
  3. Replace in the lid by tucking evenly under all lugs starting at the Gasket Offset Device. It may appear slightly wrinkled until used.
  4. Replace gasket when it begins to show signs of leakage.
  5. If a new gasket leaks, or a persistent leak develops, gently clean the sealing surface of both the lid and body of the unit with a "Scotchbrite" scrubbing pad making sure you do not remove any metal. Rinse both surfaces but do not dry.
  6. Monitor the first cycle of the day to check the Air Bleed device, which is located inside the lid, audibly "clicks" shut.
  7. Lubricate underside of body lugs with "Vaseline" if the lid becomes stiff.
  8. DO NOT LUBRICATE THE GASKET

Troubleshooting

 

Fault 1: No Power to Unit Light: L1 fails to illuminate Blown fuse / Defective Socket / Mains not connected Ensure mains lead is connected. Check / replace fuse: Check power to socket:
Fault 2: Light: L5 flashes RED Low water level or Boil dry. Allow unit to cool before refilling to the correct level. Disconnect from mains then reconnect and repeat cycle. If the fault recurs with sufficient water, arrange for a service engineer to visit.
Fault 3: Cycle failed Light: L4 fails to illuminate GREEN and there is no audible buzzer Many possible causes Disconnect from mains then reconnect and repeat cycle. If the fault recurs arrange for a service engineer to visit.
Fault 4: Steam or water leaks from under the lid I) Worn or dirty gasket. Wash gasket and sealing surfaces on the body and lid as described under "Care and Maintenance. If the fault persists, replace with a new gasket.
  ii) Incorrectly closed lid. Ensure the unit is fully depressurized by opening Depressurization Valve (A). Remove lid and refit carefully. Disconnect from mains, reconnect and repeat cycle
Fault 5: Excessive steam or water leaking from Depressurization Valve (A). Depressurization Valve (A) in "OPEN" position Close Depressurization Valve (A).

Safety Features of Retriever

In the event of a fault occurring during any stage of the units operation, identify the fault by referring to the descriptions below. The fault con be rectified by following the Fault Remedy applicable to the problem incurred.In the unlikely event that something should go wrong, we have incorporated a number of safety features to ensure that your autoclave remains safe at all times.

  1. Located to the rear of the lid, beneath the cover, is a spring called the Gasket Offset Device (GOD Spring), designed to prevent pressure building up if the lid has been incorrectly fitted. DO NOT TAMPER WITH THIS SAFETY DEVICE
  2. If for any reason, the temperature falls below the minimum required temperature, resulting in the Temperature Light (L3) switching off, the cycle timer will re-start.
  3. If there is an electrical or electronic failure resulting in a build up of pressure - in excess of normal operating pressure - one or all of the following safety features will be activated.

i) Depressurization Valve (A) will loudly and rapidly "vent" steam.

ii) The gasket will "extrude" through the slot in the rear of the lid rapidly releasing excess pressure and steam.

iii) A non - resettable thermal fuse located in the base of the unit will "melt' at a pre-determined temperature, disconnecting the power.

Should any of the devices listed activate, please observe the follwing steps:
  1. Do not touch the unit.
  2. Switch off at the wall socket and un-plug
  3. Allow temperature and pressure to drop before touching the unit and removing the content
  4. Do not attempt to re-start the unit
  5. Arrange for an immediate service by contacting PickCell Labs or your local Distributor.

Certificates of Retriever

Prestige Medical, a manufacturer of the 2100-Retriever ( belongs to the 2100 Series) has always been proud of its reputation for quality and innovation in the design and production of our sterilization systems. Prestige Medical has continually strived to be at the forefront of the industry with regards to meeting and even exceeding those standards which apply to it as a manufacturer and the products it makes:

ISO 9001 (Certificate No. FM 31988) The international standard for quality. The award of this International Standard confirms our position at the cutting edge of customer service and is the guarantee of continued quality in all areas of product design, production and servicing.

EN 46001 (Certificate No. 1354) The European quality standard for manufacturers of medical devices. EN 46001 is awarded in conjunction with ISO 9001 and is a more stringent set of standards which are applicable to manufacturers of medical devices.

MDD 93/42/EEC The European Medical Devices Directive. This directive, is mandatory for all products manufactured after the 14th June 1998, and is the new European standard to which all medical products must comply. The directive covers all aspects of product manufacturing, safety and performance.

EMC 89/366/EEC The European standard for Electro-Magnetic Compatibility. This directive applies to electrical products that are tested to ensure they do not interfere with adjacent electrical products, they are immune to electromagnetic emissions from other electrical products and they do not exceed harmful levels of electromagnetic emissions. Products that meet this directive can be 'CE' marked.

FDA Good Manufacturing Practice The award of this standard (similar to ISO 9001) by the United States Food and Drug Administration is a quality assurance system for manufacturing control and is necessary for all companies marketing medical devices in the U.S.A. (Product approval is also necessary - via 510(k)) Once awarded, the FDA conducts re-appraisals every two to three years to ensure the standards are being continually adhered to. Prestige Medical have subsequently undergone recent (March 1999) routine re-assessment to the new FDA Q.S.R. (Quality System Requirements), the latest compliance requirements.

FDA (510k) - Series 2100 FDA (510k) for medical devices awarded specifically to Prestige Medical branded Series 2100, allowing these autoclaves to be marketed in the U.S.A. This standard covers all aspects of product performance and effec2100-Retriever has been designed to comply in part or fully with the following standards:

2100-Retriever has been designed to comply in part or fully with the following standards:

IEC601: Electrical and product safety standard.

DIN 58946: Part 8. Type testing for microbiological performance.

EMC: Electro Magnetic Compatibility. EN 50081 - 1, EN 50082 - 1 and prEN 50082 - 2 are generic standards relating to EMC.

BS 3970 Part 4: British Standard for autoclaves. Pressure vessel complies.

ESCHLE: Electrical Safety Code for Hospital Laboratory Equipment (UK).

UL 544: Underwriters Laboratory (America) covering electrical and product safety.

CSA Class 8721 01: Canadian Standards Association covering electrical and product safety.

Technical Specifications

Height   335 mm
Capacity   9 litres
Max. Instrument length   228 mm
Width   340 mm
Internal chamber    
- Net Weight   4.5 kilos
- Dimensions (d/h)   210/230 mm
- Max. Load Weight.   3.0kg

Other Specifications

Fuses   Located under the control module, fuses F1 OA, 32 x 6.3mm, ceramic sand filled, Mains plug top fuse (User replaceable), F1 3A to BS1362 UK ONLY.
Rating   Models are rated continuously for intermittent use.
Body   Deep drawn aluminium.
Lid   Aluminium
Heater   Externally surface mounted mechanically fixed electric element.
Temperature Cut Out   Thermal fuse.
Pressure   Calibrated pressure release valve.
Max. Single Fault Temperature   133.3°C
Over Voltage Category   Group II Pollution Degree
Environment Conditions   - indoor use - temperature 5°C to 40°C - altitude up to 2000m - maximum relative humidity 80% for temperatures up to 31°C decreasing linearly to 50% relative humidity at 40°C. - mains supply voltage fluctuations not to exceed +10% of the nominal voltage
Input Connections   Mains inlet socket 'hot' format conforming to IEC 302.
Safety Shut Down   See “Temperature Cut Out”.
Packaging   All packaging materials are recyclable.

Technical Tip

Publications: Real-world examples of the retriever in various types of research and clinical studies.

Online Ordering

Antigen Retriever