The following materials are needed for direct eosinophil determinations:
Mixing Blood and Diluent: Immediately after drawing blood, immerse the end of the pipette into the eosinophil diluting fluid and while rotating the pipette between the fingers, draw fluid to the graduation above the bulb. Care should be taken to prevent blood from leaking out of the pipette especially when it is first put into the diluent.
Shaking the Pipette: The pipette is held between the thumb and forefinger and shaken slowly in a wide arc for approximately 10 seconds. Between each stroke, the pipette is rotated approximately 45 degrees. Care should be taken to close one or both ends of the pipette tightly with the finger to avoid leakage.
Charging the Chamber: The chambers should be loaded immediately after the pipette is shaken. Hold the pipette at a 45 degree angle, discharge 3 drops of the solution to empty the capillary stem, then place the tip of the pipette on the chamber to be loaded. Allow the fluid to run out rapidly under the cover glass and fill the chamber without overflowing at the edges. The forefinger may be placed over the rear of the pipette to control the rate of flow of the fluid.
Cells are most recognizable and easiest to count at a magnification of 150 times with a field of vision which includes a one square millimeter area. The eosinophils have a distinct cell wall enclosing numerous large red or pink granules. The remaining white blood cells are pale, usually with a wrinkled cell wall. All the red blood cells should have completely disappeared.
In order to reduce error, a systematic procedure should be followed in counting eosinophils. Numerous methods may be used. The following technique has been found to be reliable:
Each chamber consists of ten 1 square millimeter areas subdivided into 16 smaller squares (Diagram A).
As seen through the microscope, the left hand side of the chamber has an identifying number. The upper row is counted from left to right beginning with this identification number and the lower row is counted from right to left ending at the identification number (Diagram B).
Beginning in the upper left hand corner, the first vertical row of 4 squares is centered in the microscopical field. All the eosinophils within the 4 squares and those touching the ruled lines on the left and on the upper part of the squares are counted. The slide is then moved so that the next row of 4 squares is centered and the eosinophils counted in the same manner. The cells touching the rulings on the right and lower sides are not counted. The eyes can be trained to move up one row of 4 squares and down the next (Diagram C).
After completing the count, remove the cover glass and clean the counting chamber with water or a mild cleaning solution (10% solution of bleach). Dry the counting chamber with a soft cloth or wipe, or rinse with acetone.
Speirs-Levy Eosinophil Counting Chamber is available online from the EMS Catalog. For ordering or product information, click here.