UranyLess – Protocols of Use: Negative Staining
EMS Catalog #22409
Negative staining is a very useful technique in electron microscopy. It allows characterization of isolated particles of morphology as bacteria, virus, protein, nanoparticles, liposomes, exosomes, etc.
Staining Protocol #1:
- On a piece of parafilm or any other hydrophobic carrier, place a drop of your solution (~ 10µl) and a UranyLess drop.
- Using our fine tweezers, place your sample drop on a formvar-carbon coated grid. for about 1 minute.
- Blot your grid using filter paper.
- Place your grid on the UranyLess solution for 1 minute.
- Blot, let it dry for 5 minutes and observe under the microscope.

Technical Tip:
- If the staining is too intense, wash with room temperature water for 1 minute.
Staining Protocol #2 (Debra M. Townley, Baylor College of Medicine):
- One drop of suspension of 300 mesh formvar coated grids – 30 minutes
- Wick away excess suspension with paper arrow (Watman No.1)
- Place one drop of Uranyless on the grid before it is completely dry
- Stain for 3 minutes
- Wick away excess stain with paper arrow (Watman No.1)
- Allow the grids(s) to dry completely before examination on the TEM
- Times can be altered (suspension time, stain time) to suit the "look" you desire